V84FS Turbo Pump



84FS.pdf

Pumping Speed KF 40 CFF 2.75" ISO 63 / CFF 4.5”
Ar 44 l/s 57 l/s 66 l/s
N2 49 l/s 56 l/s 67 l/s
He 38 l/s 46 l/s 63 l/s
H2 36 l/s 40 l/s 53 l/s
Gas throughput at full rotational speed (with recommended forepump) Air cooling (25°C) Water cooling (25°C, 65 l/h)
N2 100 sccm 100 sccm
Ar 70 sccm 70 sccm
Compression ratio & Foreline tolerance Compression ratio Foreline tolerance
Ar > 1.0 x 1011 > 14 mbar
N2 ≥ 1.0 x 1011 > 14 mbar
He 2.0 x 106 > 12 mbar
H2 5.0 x 104 > 4 mbar
Base pressure with recommended forepump (5 m3/h) < 8.0 x 10-10 mbar (< 6.0 x 10-10  Torr)
Inlet flange   CFF 4.5”, ISO 63
CFF 2.75”, KF 40
Foreline flange KF16
Rotational speed 81,000 rpm (1350 Hz driving frequency)
Start-up time < 2 minutes
Recommended forepump   Mechanical: Agilent DS-42/102
Dry pump: Agilent SH-110/IDP-3
Operating position Any
Operating ambient temperature 5 °C to 35 °C
Relative humidity of air 0 - 90 % (not condensing)
Bakeout temperature 80°C at inlet flange max (ISO flange)
120°C at inlet flange max (CFF flange)
Lubricant Permanent lubrication
Cooling requirements Air flow temperature: (5 to 35 °C ambient)
Water (mandatory if ambient temperature > 35 °C):
Water temperature: 15 °C to 25 °C
Minimum flow: 65 l/h (0.30 GPM)
Pressure: 2 to 4 bar (45 to 75 psi)
Noise pressure level < 40 dB(A) at 1 meter, full speed
Storage temperature -40° C to 70° C
Max altitude 3000 m
Weight kg (lbs) Pump ISO 63 2.05 kg
  Pump CFF 4.5” 3.50kg
  Pump CFF 2.75” 3.34kg
  Pump KF 40” 2.37kg
Conformity to Norms EMC 61326-1
  Safety (CE/CSA) Regulatory DIR 2006/42/CE
  ROHS DIR 2011/65/EU


 

 

 

 